AbstractAbstract
[en] An alpha particle detector made of a mercuric iodide crystal grown from the vapour phase had been fabricated and tested. There was a vacuum deposition of thin film electrodes to allow alpha particles to pass through with little degradation.Test with different bias voltages to study its effect on its resolution is reported in this paper. An alpha particle detection resolution of 5% of the Am-242 had been achieved with the detector tested (author)
Primary Subject
Source
Ghana Science Association (Ghana); 112 p; Aug 2003; p. 70; 23. Biennial conference of the Ghana Science Association; Kumasi (Ghana); 4-7 Aug 2003; Available in abstract form only. Full text provided in this record; Available from the Library, Ghana Atomic Energy Commission, Box 80, Legon, Accra, Ghana