Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.018 seconds
Kessler, G.; Burns, D.T.; Buermann, L.; Prez, L.A. de
Bureau International des Poids et Mesures (BIPM), 92 - Sevres (France)2007
Bureau International des Poids et Mesures (BIPM), 92 - Sevres (France)2007
AbstractAbstract
[en] Some simulated mammography radiation beams have been established at the BIPM using a low-energy x-ray tube with a tungsten anode and molybdenum as a filter. The response of two ionization chambers of different types to these beams is compared with that obtained in mammography beams at the PTB and the NMi which were produced with x-ray tubes with molybdenum anodes and molybdenum filters. The relative differences between the chamber responses to these two different types of beams were less than 7 x 10-3 which implies the uncertainty for the transfer of a calibration from one type of beam to the other. (authors)
Primary Subject
Secondary Subject
Source
Mar 2007; 9 p; 8 refs.
Record Type
Miscellaneous
Report Number
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue