Revised comparison of the standards for air kerma of the ENEA-INMRI and the BIPM for 60Co gamma rays
Allisy-Roberts, P.J.; Burns, D.T.; Laitano, R.F.; Toni, M.; Bovi, M.
Bureau International des Poids et Mesures (BIPM), 92 - Sevres (France)2003
Bureau International des Poids et Mesures (BIPM), 92 - Sevres (France)2003
AbstractAbstract
[en] In 1998 a comparison of the standards of air kerma of the Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti of the Ente per le Nuove Tecnologie, l'Energia e l'Ambiente, Italy (ENEA-INMRI) and of the Bureau International des Poids et Mesures (BIPM) was carried out in 60Co radiation. The comparison result, declared in 2003, is 1.0044 (0.0026) and demonstrates that the ENEA-INMRI and BIPM standards are in agreement within the expanded uncertainty (k = 2). (authors)
Primary Subject
Secondary Subject
Source
Oct 2003; 12 p; 31 refs.
Record Type
Miscellaneous
Report Number
Country of publication
BETA DECAY RADIOISOTOPES, BETA-MINUS DECAY RADIOISOTOPES, CARBON, COBALT ISOTOPES, ELECTROMAGNETIC RADIATION, ELEMENTS, INTERMEDIATE MASS NUCLEI, INTERNAL CONVERSION RADIOISOTOPES, IONIZING RADIATIONS, ISOMERIC TRANSITION ISOTOPES, ISOTOPES, MATHEMATICAL SOLUTIONS, MEASURING INSTRUMENTS, MINERALS, MINUTES LIVING RADIOISOTOPES, NONMETALS, NUCLEI, NUMERICAL SOLUTION, ODD-ODD NUCLEI, RADIATION DETECTORS, RADIATIONS, RADIOISOTOPES, STANDARDS, YEARS LIVING RADIOISOTOPES