Wunderlich, Adam; Noo, Frederic, E-mail: awunder@ucair.med.utah.edu, E-mail: noo@ucair.med.utah.edu2008
AbstractAbstract
[en] We consider noise in computed tomography images that are reconstructed using the classical direct fan-beam filtered backprojection algorithm, from both full- and short-scan data. A new, accurate method for computing image covariance is presented. The utility of the new covariance method is demonstrated by its application to the implementation of a channelized Hotelling observer for a lesion detection task. Results from the new covariance method and its application to the channelized Hotelling observer are compared with results from Monte Carlo simulations. In addition, the impact of a bowtie filter and x-ray tube current modulation on reconstruction noise and lesion detectability are explored for full-scan reconstruction
Primary Subject
Source
S0031-9155(08)64305-5; Available from http://dx.doi.org/10.1088/0031-9155/53/10/002; Country of input: International Atomic Energy Agency (IAEA)
Record Type
Journal Article
Journal
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue