Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.017 seconds
Ku, Benjamin Y.; Chan, M.-L.; Ma Zhiya; Horsley, David A., E-mail: dahorsley@ucdavis.edu2008
AbstractAbstract
[en] Optical detection of the frequency-dependent magnetic relaxation signal is used to monitor the binding of biological molecules to magnetic nanoparticles in a ferrofluid. Biological binding reactions cause changes in the magnetic relaxation signal due to an increase in the average hydrodynamic diameter of the nanoparticles. To allow the relaxation signal to be detected in dilute ferrofluids, measurements are made using a balanced photodetector, resulting in a 25 μV/√Hz noise floor, within 50% of the theoretical limit imposed by photon shot noise. Measurements of a ferrofluid composed of magnetite nanoparticles coated with anti-IgG antibodies show that the average hydrodynamic diameter increases from 115.2 to 125.4 nm after reaction with IgG
Primary Subject
Source
S0304-8853(08)00586-6; Available from http://dx.doi.org/10.1016/j.jmmm.2008.04.132; Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA)
Record Type
Journal Article
Journal
Journal of Magnetism and Magnetic Materials; ISSN 0304-8853;
; CODEN JMMMDC; v. 320(18); p. 2279-2283

Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue