Pau, Stanley; Whitten, William B.; Ramsey, Dr. John Michael
Oak Ridge National Laboratory (United States). Funding organisation: ORNL work for others (United States)2007
Oak Ridge National Laboratory (United States). Funding organisation: ORNL work for others (United States)2007
AbstractAbstract
[en] A planar quadrupole ion trap is proposed. We have demonstrated an extremely large operating range by trapping ions and particles with mass-to-charge ratio ranging from 102 to 109 at frequencies from 2.8 x 106 to 60 Hz at an operating pressure of 1.1 x 10-4 to 760 Torr, respectively, using a trap radius of r1 = 1 mm. We have also performed mass spectrometry with a resolution of 1.2 amu with mass-to-charge range from 50 to 150. Our geometry is simple enough to be integrated into existing integrated circuits and microelectromechanical system devices, opening up the possibility of many novel hybrid applications and experiments
Primary Subject
Source
ORNL/PTS--2645; 400403809; AC05-00OR22725
Record Type
Journal Article
Journal
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue