Hansen, K.; Reckleben, C.; Diehl, I.; Klaer, H., E-mail: karsten.hansen@desy.de2008
AbstractAbstract
[en] A new Si-drift detector module for fast X-ray spectroscopy experiments was developed and realized. The Peltier-cooled module comprises a sensor with 7x7-mm2 active area, an integrated circuit for amplification, shaping and detection, storage, and derandomized readout of signal pulses in parallel, and amplifiers for line driving. The compactness and hexagonal shape of the module with a wrench size of 16 mm allow very short distances to the specimen and multi-module arrangements. The power dissipation is 186 mW. At a shaper peaking time of 190 ns and an integration time of 450 ns an electronic rms noise of ∼11 electrons was achieved. When operated at 70C, FWHM line widths around 260 and 460 eV (Cu-Kα) were obtained at low rates and at sum-count rates of 1.7 MHz, respectively. The peak shift is below 1% for a broad range of count rates. At 1.7-MHz sum-count rate the throughput loss amounts to 30%
Primary Subject
Source
S0168-9002(08)00214-3; Available from http://dx.doi.org/10.1016/j.nima.2008.02.010; Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA)
Record Type
Journal Article
Journal
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment; ISSN 0168-9002;
; CODEN NIMAER; v. 589(2); p. 250-258

Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue