Smith, M.L.; Bignell, L.; Alexiev, D.; Mo, L.; Harrison, J., E-mail: michael.smith@ansto.gov.au2008
AbstractAbstract
[en] Three types of lead shielding assemblies were investigated to establish which provides the lowest background as preliminary work for determining the efficiency of a High Purity Germanium (HPGe) detector. The background continuum and peaks were examined using gamma spectroscopy with a coaxial and a planar HPGe detector. The 210Pb concentration in each lead sample taken from the assemblies was quantitatively determined. A French lead sample was found to have the lowest background in this relative comparison with a 210Pb concentration of 29.1±1.3 Bq kg-1
Primary Subject
Source
S0168-9002(08)00255-6; Available from http://dx.doi.org/10.1016/j.nima.2008.02.050; Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA)
Record Type
Journal Article
Journal
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment; ISSN 0168-9002;
; CODEN NIMAER; v. 589(2); p. 275-279

Country of publication
ALPHA DECAY RADIOISOTOPES, BETA DECAY RADIOISOTOPES, BETA-MINUS DECAY RADIOISOTOPES, ELEMENTS, EVALUATION, EVEN-EVEN NUCLEI, GE SEMICONDUCTOR DETECTORS, HEAVY NUCLEI, ISOTOPES, LEAD ISOTOPES, MEASURING INSTRUMENTS, METALS, NUCLEI, RADIATION DETECTORS, RADIOISOTOPES, SEMICONDUCTOR DETECTORS, SPECTROSCOPY, YEARS LIVING RADIOISOTOPES