Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.019 seconds
AbstractAbstract
[en] We developed a simple geometric calibration method for modified animal pinhole single photon emission computed tomography (SPECT) imaging systems using a laser alignment system that does not require additional calibration scans, and verified the feasibility of this system. An optical system consisting of a laser generator and a reflecting mirror, and an alignment method using this system were developed. After laser alignment had been completed, SPECT scans of a Tc-99m line source were performed. After acquiring data with complete alignment, the rotation stage was moved along the y-axis (parallel to the detector plane) to acquire off-axis data. Using these on- and off-axis settings, a micro-performance phantom with hot-rod inserts was scanned after filling with Tc-99m solution. Reconstructed images of the line source and the hot-rod phantom showed that image degradation was minimized when the rotation center was completely aligned using our system. The spatial resolution of the reconstructed image measured using the line source was finest under complete alignment. Under this condition, the smallest hot rod (1.2 mm diameter) was resolved in the SPECT image acquired using a 0.5 mm pinhole aperture. The effects of misalignment were clearly observable in off-axis images; only hot rods with more than 3.2 mm diameter were resolved in 1 mm off-axis image whereas there were no resolvable hot rods with more than 2 mm off-axis image. This system of which the feasibility was verified in the present study will be useful for low-cost molecular imaging studies using single photon emitting tracers
Primary Subject
Source
S0168-9002(08)00205-2; Available from http://dx.doi.org/10.1016/j.nima.2008.02.033; Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA)
Record Type
Journal Article
Journal
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment; ISSN 0168-9002;
; CODEN NIMAER; v. 589(2); p. 338-344

Country of publication
BETA DECAY RADIOISOTOPES, BETA-MINUS DECAY RADIOISOTOPES, BOSONS, COMPUTERIZED TOMOGRAPHY, DIAGNOSTIC TECHNIQUES, DISPERSIONS, ELEMENTARY PARTICLES, EMISSION COMPUTED TOMOGRAPHY, HOMOGENEOUS MIXTURES, HOURS LIVING RADIOISOTOPES, INTERMEDIATE MASS NUCLEI, INTERNAL CONVERSION RADIOISOTOPES, ISOMERIC TRANSITION ISOTOPES, ISOTOPES, MASSLESS PARTICLES, MIXTURES, MOCKUP, MOTION, NUCLEI, ODD-EVEN NUCLEI, OPENINGS, RADIOISOTOPES, RESOLUTION, STRUCTURAL MODELS, TECHNETIUM ISOTOPES, TOMOGRAPHY, YEARS LIVING RADIOISOTOPES
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue