Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.025 seconds
AbstractAbstract
[en] In this work radiofrequency-coils for the imaging of small animals in clinical whole-body MRI-systems were developed. Therefore in a first step single-channel solenoids were designed and characterized. The solenoids had two and three windings respectively, which were implemented as double wires to increase the homogeneity of the receive profile. These coils allow the acquisition of whole-body images of mice with high signal-to-noise ratio and homogeneity over a distance of at least 6.3 cm. Since many imaging experiments require rapid image acquisition, in the next step a novel coil concept was developed, which, due to its geometry, enables parallel imaging in arbitrary directions. A prototype was assembled and tested on phantom and small-animal experiments. With an accelerating factor of R=2, the difference of the SNR in all directions from the theoretical maximum, was less than 1%. In order to compensate physiological motion by the self-gating technique, in this work a coil is presented for the first time, which selectively amplifies the self-gating signal, while - due to a optical detuning technique - preserving the homogeneous illumination of the image. In vivo experiments on a small animal show an amplification of the self-gating signal by at least 40%. (orig.)
Original Title
Spulenkonzepte zur schnellen und bewegungskompensierten MR-Bildgebung von Kleintieren
Primary Subject
Source
6 May 2009; 137 p; Diss.
Record Type
Miscellaneous
Literature Type
Thesis/Dissertation
Report Number
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue