AbstractAbstract
[en] In this study it was carried out the development and implementation of a system for the appropriate process of gamma radiation monitors calibration, constituted by a pneumatic dispositive to exchange the attenuators and a positioning table, both actuated through a control panel. We also implemented a System of Caesa-Gammatron Irradiator, which increased the range of the air kerma rates, due to its higher activity comparing with the current system of gamma radiation in use in the calibration laboratory of gamma irradiation. Hence, it was necessary the installation of an attenuator dispositive remotely controlled in this irradiator system. Lastly, it was carried out an evaluation of the reduction in the rates of the occupational dose. This dissertation was developed with the aim of improving the quality of the services of calibration and tests of gamma radiation monitors - provided by the IPEN Laboratory of Instrument Calibration - as well as decreasing the occupational dose of the technicians involved in the process of calibration, following thus the principles of radiation protection. (author)
Original Title
Desenvolvimento e implantacao de um sistema automatizado para adequacao do processo de calibracao de monitores de radiacao gama
Primary Subject
Secondary Subject
Source
2012; 95 p; Diss. (M.Sc.)
Record Type
Miscellaneous
Literature Type
Thesis/Dissertation
Report Number
Country of publication
LanguageLanguage
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue