AbstractAbstract
[en] Stacks composed by 1 mm thickness CR-39 foils and polyethylene and PVC films were evaluated to be used as neutron dosemeters. Irradiations were made with a calibrated 241Am-Be source in a dose range from 0 to 3.1 mSv and the etching conditions were optimized. The measurements of number of tracks per surface unit in the CR-39 detectors showed a good linear behaviour as a function of the dose. The minimum detectable dose equivalent (MDDE) was calculated. (author)
Original Title
Dosimetria de neutrones utilizando CR-39 como detector de trazas nucleares : calibracion del sistema
Primary Subject
Source
2009; 5 p; AATN; Buenos Aires (Argentina); AATN 2009: 36. Annual meeting of the Argentine Association of Nuclear Technology: Nuclear energy in the present world; AATN 2009: 36. Reunion anual de la Asociacion Argentina de Tecnologia Nuclear (AATN): La energia nuclear en el mundo actual; Buenos Aires (Argentina); 16-20 Nov 2009; 7 refs., 3 figs.