AbstractAbstract
[en] Etched-track polyallyl diglycol carbonate (PADC) dosemeters have been in use at the Paul Scherrer Institute since 1998 in neutron dosimetry for individual monitoring. In the last years, the availability of PADC materials from different manufacturers has grown, and different etching conditions were proposed, with the intention to improve the quality and overall performance of PADC in individual neutron monitoring. The goal of the present study was to compare the performance of different PADC materials and to investigate the influence of different etching conditions on sensitivity to fast neutrons and lower detection limit. The comparison covers six different PADC materials and eight different etching conditions. (authors)
Primary Subject
Secondary Subject
Source
IM2015: 5. International Conference on Individual Monitoring of Ionising Radiation; Bruges (Belgium); 20-24 Apr 2015; Available from doi: http://dx.doi.org/10.1093/rpd/ncv421; Country of input: France; 17 refs.
Record Type
Journal Article
Literature Type
Conference
Journal
Radiation Protection Dosimetry; ISSN 0144-8420;
; v. 170(1-4); p. 162-167

Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue
External URLExternal URL