AbstractAbstract
[en] In this work, we are analysing the dose response of HEMAGIC polymer gel dosimeter. Radiation sensitive polymer gel is prepared according to the literature. Gel is poured into the plastic cuvette and covered with aluminium foil in order to avoid autopolymerisation then kept in 4°C for gelling purpose. Dosimetric cuvette is irradiated with 6 MV X-rays from the medical linear accelerator with dose ranging from 2 to 14 Gy at 5 cm depth. Powerful double beam UV-visible spectrophotometer is used to study the optical absorbance
Primary Subject
Source
ICRR-HHE 2016: international conference on radiation research - impact on human health and environment and first biennial meeting of society for radiation research: abstracts; Mumbai (India); 11-13 Oct 2016
Record Type
Journal Article
Literature Type
Conference
Journal
Journal of Radiation and Cancer Research; ISSN 0973-0168;
; v. 7(1); p. 13

Country of publication
BETA DECAY RADIOISOTOPES, BETA-MINUS DECAY RADIOISOTOPES, CHEMICAL DOSEMETERS, COBALT ISOTOPES, DIAGNOSTIC TECHNIQUES, DOSEMETERS, INTERMEDIATE MASS NUCLEI, INTERNAL CONVERSION RADIOISOTOPES, ISOMERIC TRANSITION ISOTOPES, ISOTOPES, MEASURING INSTRUMENTS, MEDICINE, MINUTES LIVING RADIOISOTOPES, NUCLEAR MEDICINE, NUCLEI, ODD-ODD NUCLEI, RADIOISOTOPES, RADIOLOGY, THERAPY, TOMOGRAPHY, YEARS LIVING RADIOISOTOPES