Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.015 seconds
Jung, Jiwoong; Choi, Yong; Jung, Jin Ho; Kim, Sangsu; Im, Ki Chun, E-mail: ychoi@sogang.ac.kr, E-mail: jinho1115@gmail.com2016
AbstractAbstract
[en] Recently, we have developed the second prototype Silicon photomultiplier (SiPM) based positron emission tomography (PET) scanner for human brain imaging. The PET system was comprised of detector block which consisted of 4×4 SiPMs and 4×4 Lutetium Yttrium Orthosilicate arrays, charge signal transmission method, high density position decoder circuit and FPGA-embedded ADC boards. The purpose of this study was to evaluate the performance of the newly developed neuro-PET system. The energy resolution, timing resolution, spatial resolution, sensitivity, stability of the photo-peak position and count rate performance were measured. Tomographic image of 3D Hoffman brain phantom was also acquired to evaluate imaging capability of the neuro-PET. The average energy and timing resolutions measured for 511 keV gamma rays were 17±0.1% and 3±0.3 ns, respectively. Spatial resolution and sensitivity at the center of field of view (FOV) were 3.1 mm and 0.8%, respectively. The average scatter fraction was 0.4 with an energy window of 350–650 keV. The maximum true count rate and maximum NECR were measured as 43.3 kcps and 6.5 kcps at an activity concentration of 16.7 kBq/ml and 5.5 kBq/ml, respectively. Long-term stability results show that there was no significant change in the photo-peak position, energy resolution and count rate for 60 days. Phantom imaging studies were performed and they demonstrated the feasibility for high quality brain imaging. The performance tests and imaging results indicate that the newly developed PET is useful for brain imaging studies, if the axial FOV is extended to improve the system sensitivity.
Primary Subject
Secondary Subject
Source
S0168-9002(16)30022-5; Available from http://dx.doi.org/10.1016/j.nima.2016.03.004; Copyright (c) 2016 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA)
Record Type
Journal Article
Journal
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment; ISSN 0168-9002;
; CODEN NIMAER; v. 819; p. 182-187

Country of publication
BODY, CENTRAL NERVOUS SYSTEM, COMPUTERIZED TOMOGRAPHY, DIAGNOSTIC TECHNIQUES, ELECTROMAGNETIC RADIATION, ELECTRONIC EQUIPMENT, EMISSION COMPUTED TOMOGRAPHY, ENERGY RANGE, EQUIPMENT, IONIZING RADIATIONS, KEV RANGE, MOCKUP, NERVOUS SYSTEM, ORGANS, PHOTOTUBES, RADIATIONS, RESOLUTION, STRUCTURAL MODELS, TOMOGRAPHY
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue