Lazzaro, M.; Friedrich, B.Q.; Luz, R.M. da; Silva, A.M.M. da, E-mail: marcos.lazzaro@acad.pucrs.br
Associacao Brasileira de Fisica Medica, Rio de Janeiro, RJ (Brazil)2016
Associacao Brasileira de Fisica Medica, Rio de Janeiro, RJ (Brazil)2016
AbstractAbstract
[en] With the development of new technologies, have emerged new conversion methods of X-ray image, such as flat panel detectors. The aim of this work is the comparison of entrance surface air kerma (ESAK) and image quality between an image intensifier type of detector (A) and a flat panel (B). The ESAK was obtained by placing a ionization chamber under PMMA simulators of 10, 20 and 30 cm and the image quality was obtained by using the TOR "1"8FG simulator. The ESAK to the equipment A is higher when compared to the equipment B. The high contrast resolution is better for the equipment A for all thicknesses of simulators. The equipment A has low contrast resolution with a better viewing threshold for thicknesses of 10 and 20 cm, and a worse performance for 30 cm. It is concluded that the equipment B has ESAK smaller and despite having lower resolution, in almost all cases, have appropriate image quality for diagnosis. (author)
Original Title
Comparacao da dose e qualidade da imagem de um detector Flatpanel e um intensificador de imagem
Primary Subject
Source
2016; 4 p; 21. Brazilian congress on medical physics; Florianopolis, SC (Brazil); 24-27 Aug 2016
Record Type
Miscellaneous
Literature Type
Conference
Report Number
Country of publication
LanguageLanguage
Reference NumberReference Number
INIS VolumeINIS Volume