AbstractAbstract
[en] A signal tapping circuit has been designed to derive the fast timing signal from the Chevron type, 2 stage MCP having phosphor screen as a readout device. Using this, the timing property of the MCP-BaF2 system has been studied using the digital storage oscilloscope in which a timing resolution of ∼ 500 ps has been obtained. The applicability of MCP with phosphor screen configuration for timing measurements has been demonstrated. (author)
Primary Subject
Source
14. biennial symposium on Nuclear and Radiochemistry (NUCAR-2019); Mumbai (India); 15-19 Jan 2019; 11 refs.
Record Type
Journal Article
Literature Type
Conference
Journal
Journal of Radioanalytical and Nuclear Chemistry; ISSN 0236-5731;
; CODEN JRNCDM; v. 323(3); p. 1419-1424