[en] We have developed and tested a thermal neutron, two-dimensional position-sensitive detector for use in neutron scattering from biological samples. The detector is based on a multi-wire proportional concept with a gas filling of 6 atmos. 3He, 4 atmos. A, and 0.5 atmos. CO2. The position readout uses a resistive wire with charge division to determine position. We have measured a resolution of less than 3 mm (FWHM) with a detection efficiency of 70 percent at 2.8 A neutrons. Considerations leading to an optimal configuration are discussed. (U.S.)
Source
21. nuclear science symposium; Washington, DC; 11 Dec 1974; 14. scintillation and semiconductor counter symposium; Washington, DC; 11 Dec 1974; 6. nuclear power systems symposium; Washington, DC; 11 Dec 1974
Record Type
Journal Article
Literature Type
Conference
Journal
IEEE (Inst. Electr. Electron. Eng.) Trans. Nucl. Sci; v. NS-22(1); p. 255-268
Country of publication
BARYONS, CARBON COMPOUNDS, CARBON OXIDES, CHALCOGENIDES, DISPERSIONS, ELEMENTARY PARTICLES, ELEMENTS, EVEN-ODD NUCLEI, FERMIONS, HADRONS, HELIUM ISOTOPES, ISOTOPES, LIGHT NUCLEI, MEASURING INSTRUMENTS, NEUTRONS, NONMETALS, NUCLEI, NUCLEONS, OXIDES, OXYGEN COMPOUNDS, PROPORTIONAL COUNTERS, RADIATION DETECTION, RADIATION DETECTORS, RARE GASES, RESOLUTION, STABLE ISOTOPES