AbstractAbstract
[en] A method of measuring μtau and lambda of electrons and holes in a semi-insulating CdTe single crystal is described. The values were obtained from alpha spectra measurements with a detector using the CdTe single crystal. The obtained parameters were introduced into a relation for gamma spectra calculation; the calculated spectra agreed very well with experimental results. (author)
Source
Dec 1975; 14 p; Available from Ustav Jaderneho Vyzkumu, Prague-Rez, Czechoslovakia.