Kamboj, B.K.; Shahnai, M.G.; Phadnis, U.V.; Sharma, D.
Bhabha Atomic Research Centre, Bombay (India)1976
Bhabha Atomic Research Centre, Bombay (India)1976
AbstractAbstract
[en] The report describes a simple and reproducible 4π polythene geometry incorporating He3 detector for the calibration of neutron sources. The response of the system has been shown to be constant in the energy range of 2.1 MeV to 4.5 MeV, using radioactive neutron sources standardized with an improved manganese bath technique. A wide range of neutron sources having emission rates in the range 103 to 106 neutrons per/sec. can be calibrated using this detector with an accuracy of about 6%. It is expected that like the Precision Long Counter, this detector could serve as a secondary standard for the calibration of neutron sources. (author)
Source
1976; 5 p; 4 figures.