Published 2017
| Version v1
Miscellaneous
A comparison of Cs-137 and x-ray sources as calibration references for thermoluminescent dosimeter chips used in the Saskatchewan postal quality assurance program for diagnostic radiology
Creators
- 1. Univ. of Saskatchewan, Saskatoon, SK (Canada)
- 2. Radiation Safety Inst. of Canada, Saskatoon, SK (Canada)
Description
This work is focused on calibrating Thermoluminescent Dosimeters (TLDs) using Cs-137 instead of using x-ray. TLD chips which are used in the Saskatchewan Ministry of Labour Postal Quality Assurance (PQA) program were irradiated using X-Ray machine and an AEA Model 773 Cs-137 Instrument Calibrator at known exposure levels. The response was measured to be 0.1275 ± 0.0049 nC per mR for x-ray exposures and 0.0772 ± 0.0028 nC per mR for the Cs-137 exposures. Neither kVp or beam filtration appeared to have any effect on the chip response. These results show that application of an appropriate correction factor k =0.6056 ± 0.0057 will allow chips to be calibrated using Cs-137. (author)
Availability note (English)
Available as a slide presentation alsoAdditional details
Publishing Information
- Publisher
- Canadian Nuclear Society
- Imprint Place
- Toronto, Ontario (Canada)
- ISBN
- 978-1-926773-24-7
- Imprint Title
- Our nuclear future: renewal and responsibility. 37th Annual CNS conference and 41st CNS/CNA student conference
- Imprint Pagination
- 408 Megabytes
- Journal Page Range
- [5 p.]
Conference
- Title
- 37. annual Canadian Nuclear Society conference; 41. CNS/CNA student conference
- Dates
- 4-7 Jun 2017
- Place
- Niagara Falls, Ontario (Canada)
INIS
- Country of Publication
- Canada
- Country of Input or Organization
- Canada
- INIS RN
- 50009050
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference, Non-conventional Literature
- Descriptors DEI
- CALIBRATION STANDARDS; CESIUM 137; QUALITY ASSURANCE; RADIATION DOSES; THERMOLUMINESCENT DOSEMETERS
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CESIUM ISOTOPES; DOSEMETERS; DOSES; INTERMEDIATE MASS NUCLEI; ISOTOPES; LUMINESCENT DOSEMETERS; MANAGEMENT; MEASURING INSTRUMENTS; NUCLEI; ODD-EVEN NUCLEI; QUALITY MANAGEMENT; RADIOISOTOPES; STANDARDS; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- 7 refs., 3 figs.